Main characteristics
Starting voltage |
V |
260÷320 |
Operating voltage (recommended) |
V |
400 |
Operating voltage range |
V |
350÷450 |
Plateau length |
V |
100 min |
Plateau slope |
%·V-1 |
0,2 max |
Sensitivity to 137Cs with dose rate 2 µR·s-1 |
µR-1 |
136÷184 |
Own background in the 50 mm Pb shield |
cps |
1 max |
Operating temperature range |
°С |
-50÷+60 |
Life expectancy |
count |
1×1010 |
Weight |
g |
20 max |