Helium-18/190-8,0/OC

Download Specifications in .pdf format

3Не – neutron counter

Neutron counter Helium-18/190-8,0/OC


Main characteristics

Operating mode

proportional

Effective length

mm

133,5

Pressure of filling gas mixture atm 8,0
3Не partial pressure atm 8,0
Operating voltage (recommended) V 1400
Maximum permissible operating voltage V 1700
Peak width at half height % 15 max
Variation of the position of the peak of the mean value in a party % 20 max
Own background count / s 0,1 max
The variation of sensitivity relative to the control sample % 5 max
Operating temperature range °C -50 ÷ 60
Insulation resistance Ω 1×1012 min
Weight

g

51

Overall dimensions

neutron counter overall dimensions

Pulse-height spectrum on the recommended operating voltage

neutron counter spectrum

on top

Дополнительная информация