Main characteristics
Starting voltage |
V |
410 max |
Working voltage |
V |
550 |
Plateau length |
V |
500 ÷ 600 |
Plateau slope |
% / V |
0,3 max |
Sensitivity to 137 Cs |
(count/s)/ (mR/h) |
0,32÷0,38 |
Own background (in the 50 mm Pb shield) |
count /s |
0,02 max |
Energy Compensation Range | keV | 80 ÷ 1250 |
Sensitivity change relative to 137Cs within the range of energy compensation | % | -15 ÷ 15 |
Life expectancy |
count |
1×1010 |
Operating temperature range | °С | -40 ÷ +70 |
Weight |
g |